Remote control commands
R&S
®
RTO6
1899User Manual 1801.6687.02 ─ 05
Query parameters:
<SearchName> String parameter that contains the search definition name
Return values:
<Translation> Symbolic label (string)
Usage: Query only
24.17.11 SENT (option R&S RTO6-K520)
● Configuration.......................................................................................................1899
● T
rigger.................................................................................................................1902
● Decode results.................................................................................................... 1909
● SENT search settings......................................................................................... 1916
● SENT search results........................................................................................... 1924
24.17.11.1 Configuration
BUS<m>:SENT:DATA:SOURce..................................................................................... 1899
BUS<m>:SENT:DATA:THReshold..................................................................................1899
BUS<m>:SENT:TECHnology.........................................................................................1900
BUS<m>:SENT:CLKPeriod........................................................................................... 1900
BUS<m>:SENT:CLKTolerance.......................................................................................1900
BUS<m>:SENT:DNIBbles............................................................................................. 1901
BUS<m>:SENT:SFORmat.............................................................................................1901
BUS<m>:SENT:CRCVersion......................................................................................... 1901
BUS<m>:SENT:CRCMethod......................................................................................... 1901
BUS<m>:SENT:PPULse............................................................................................... 1902
BUS<m>:SENT:PPFLength...........................................................................................1902
BUS<m>:SENT:DATA:SOURce <DataSource>
Selects the source of the data line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 |
R1 | R2 | R3 |
R4
*RST: C1W1
Usage: Asynchronous command
BUS<m>:SENT:DATA:THReshold <Threshold>
Sets a user-defined threshold value. Alternatively, you can set the threshold according
to the signal technology BUS<m>:
SENT:TECHnology.