TRIGger<m>:LEVel<n>:NOISe:PERDivision <HystInDivs>
Defines a range in divisions around the trigger level. If the signal oscillates inside this
range and crosses the trigger level thereby
, no trigger event occurs.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
<n> 1..31
Indicates the trigger source: see TRIGger<m>:LEVel<n>:
NOISe[:
STATe] on page 1424.
Parameters:
<HystInDivs> Hysteresis size in divisions
Range: 0 to 5
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:NOISe:RELative <HystRel>
Defines a range in percent around the trigger level. If the signal oscillates inside this
range and crosses the trigger level thereby, no trigger event occurs.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
<n> 1..31
Indicates the trigger source: see TRIGger<m>:LEVel<n>:
NOISe[:
STATe] on page 1424.
Parameters:
<HystRel> Hysteresis in %. 10% = 1 div
Range: 0 to 50
Increment: 1
*RST: 0
Default unit: %
Usage: Asynchronous command
24.9.18 Trigger sequence
TRIGger<m>:
EVENt[:EVENt]........................................................................................ 1427
TRIGger<m>:SEQuence:TYPE..................................................................................... 1427
TRIGger<m>:ECOupling...............................................................................................1428
TRIGger<m>:SEQuence:DELay.................................................................................... 1428
TRIGger<m>:SEQuence:COUNt................................................................................... 1428
Trigger

Table of Contents

Other manuals for R&S RTO6

Related product manuals