EasyManua.ls Logo

R&S RTO6 - Page 1428

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1428User Manual 1801.6687.02 ─ 05
TRIGger<m>:ECOupling <T
rigLevEvtCpl>
Event coupling of trigger levels: Sets the trigger levels to the values of the indicated
event. Thus, channel 1 has one trigger level for all events, channel 2 has one trigger
level and so on.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<TrigLevEvtCpl> ON | OFF
*RST: ON
Usage: Asynchronous command
TRIGger<m>:SEQuence:DELay <Delay>
Sets the time the instrument waits after an A-event until it recognizes B-events.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Delay> Range:
0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
TRIGger<m>:SEQuence:COUNt <Events>
Sets the number of B-events to be fulfilled after an A-event. The last B-event causes
the trigger.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Events> Range:
1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command
TRIGger<m>:SEQuence:RESet:EVENt <EnabRstEvt>
If set to ON, the trigger sequence is restarted by the R-event if the specified number of
B-event does not occur.
Trigger