Jitter analysis and clock data recovery
R&S
®
RTO6
1119User Manual 1801.6687.02 ─ 05
Trigger level
Sets the voltage level for the trigger event. For CDR, the trigger level is also the thresh-
old level for the specified input channel.
Remote command:
TRIGger<m>:LEVel<n>[:
VALue] on page 1391
18.3.2.3 Serial pattern trigger using CDR
If option R&S RTO6-K13 is installed, you can use the recovered clock signal for the
serial pattern trigger instead of a real clock signal.
1. Select the trigger source and the trigger type "Serial pattern".
2. Set "Clock / CDR" to "CDR".
3. Tap "Hardware CDR" to set up the clock data recovery. See Chapter 18.3.2.1,
"Hardware CDR setup", on page 1116 for details.
Clock data recovery