Remote control commands
R&S
®
RTO6
2435User Manual 1801.6687.02 ─ 05
The effect depends on TRIGger<m>:
PARallel:
WIDTh:RANGe.
●
For the ranges SHORter and LONGer, the width defines the maximum and mini-
mum pulse width, respectively.
●
For the ranges WITHin and OUTSide, the width defines the center of a range
which is defined by the limits TRIGger<m>:PARallel:WIDTh:DELTa.
Parameters:
<Width> Range: 200E-12 to 10000
Increment: 200E-9
*RST: 5E-9
Default unit: s
Usage: Asynchronous command
TRIGger<m>:PARallel:WIDTh:DELTa <WidthDelta>
Defines a range around the given width value. the setting is relevant if TRIGger<m>:
PARallel:
WIDTh:RANGe is set to WITHin or OUTSide. The width is set with
TRIGger<m>:
PARallel:WIDTh:WIDTh.
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 600E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
24.18.4.4 Timeout trigger
See also:
●
TRIGger<m>:
PARallel:TIMeout:EXPRession[:DEFine] on page 2433
TRIGger<m>:PARallel:TIMeout:RANGe......................................................................... 2435
TRIGger<m>:PARallel:TIMeout:TIME.............................................................................2436
TRIGger<m>:PARallel:TIMeout:RANGe <TimeoutMode>
Sets the state condition.
Parameters:
<TimeoutMode> HIGH | LOW | EITHer
HIGH
The level of a digital channel stays above the threshold, or the
logical expression for LOGic trigger source is true.
LOW
The level of a digital channel stays below the threshold, or the
logical expression for LOGic trigger source is false.
EITHer
The signal state remains unchanged.
Mixed signal option (MSO, R&S RT
O6-B1)