Mask testing
R&S
®
RTO6
405User Manual 1801.6687.02 ─ 05
Figure 10-5: Waveform mask with vertical width = 0.5 div and vertical position = -0.5 div
Remote command:
MTESt:WFMRescale:
YPOSition on page 1594
10.2.3 Define Test
Access: [App Cockpit] > "Analysis" tab > "Mask" > "Define Test"
There are two usual test practices:
Testing a defined number of waveforms against the mask and initiate an action
when the acquisition cycle has been completed without failure:
Set the number of acquisitions to be tested: "Average count (N-single count)"
Start [Single]
Testing a continuous acquisition or a defined number of waveforms against the
mask and initiate an action when the fail criteria is fulfilled.
Make sure that the correct mask test tab is selected before you enter the settings.
Fail condition, Violation tolerance
The fail criteria for a mask test is set by two parameters: "Fail condition" and "V
iolation
tolerance".
"Fail condition" defines the kind of hits to be considered for test evaluation:
"Samples": Considers the number of samples that hit the mask.