EasyManua.ls Logo

R&S RTO6 - 19.3.1.2 Configuring power quality

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1140User Manual 1801.6687.02 ─ 05
POWer:QUALity:RESult:
POWer:PHASe? on page 2528
POWer:QUALity:RESult:POWer:REACtive? on page 2528
POWer:QUALity:RESult:POWer:REALpower? on page 2528
POWer:QUALity:RESult:VOLTage:CREStfactor? on page 2528
POWer:QUALity:RESult:VOLTage:FREQuency? on page 2528
POWer:QUALity:RESult:VOLTage:PEAK? on page 2528
POWer:QUALity:RESult:VOLTage:RMS? on page 2528
POWer:QUALity:REPort:ADD on page 2528
19.3.1.2 Configuring power quality
For details of the configuration settings, see Chapter 19.3.1.3, "Power quality settings",
on page 1141.
1. Select "Analysis" > "Power".
2. Under "Input", select "Power Quality".
3. Connect the differential voltage probe and the current probe to the oscilloscope.
4. Deskew the probes as described in Chapter 19.1.1.1, "Auto deskew",
on page 1123.
5. Connect the probes to the DUT as shown in the "Channels" tab:
Input

Table of Contents

Other manuals for R&S RTO6

Related product manuals