Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1141User Manual 1801.6687.02 ─ 05
6. Select the correct channels of the "Current Source" and the "Voltage Source".
7.
Select the "Settings" tab.
8.
Set the "Input frequency" according to your signal.
9. Select an "Optimize Scaling" option.
10. Tap "Execute".
On the screen you can see the measurement waveforms of the current, the voltage
and the power. Also, the result box with numeric measurement results is shown.
For details, see Chapter 19.3.1.1, "Power quality results", on page 1138.
19.3.1.3 Power quality settings
In the "Channels" tab, you set the current source and the voltage source, see also:
Chapter 19.2.1, "Channels tab", on page 1134.
In the "Settings" tab, you configure the power measurement parameters and display
settings.
Input frequency
Selects the input frequency of the source signal.
Remote command:
POWer:QUALity:FREQ
on page 2527
Input