R&S RTO6 - 13.8.2 LIN configuration; 13.8.2.1 LIN configuration settings

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTO6
606User Manual 1801.6687.02 ─ 05
13.8.2 LIN configuration
13.8.2.1 LIN configuration settings
Access: [Bus] >
"LIN"
> "Setup" tab
Make sure that the tab of the correct serial bus is selected.
See also: Chapter 13.1.1, "Setup - general settings", on page 485
LIN standard
Selects the version of the LIN standard that is used in the DUT. The setting mainly
defines the checksum version used during decoding.
The most common version is LIN 2.x. For mixed networks, or if the standard is
unknown, set the LIN standard to
"Auto".
Remote command:
BUS<m>:LIN:STANdard on page 1839
Data
Sets the source waveform of the data line.
LIN (option R&S RT
O6-K520)

Table of Contents