Remote control commands
R&S
®
RTO6
1417User Manual 1801.6687.02 ─ 05
TRIGger<m>:TV:CUSTom:SDURation <SyncPulseDurat>
Sets the width of the sync pulse. Only relevant if TRIGger<m>:
TV:
STANdard is set to
CUSTom.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<SyncPulseDurat>
Range: 100E-9 to 100E-6
Increment: 100E-9
*RST: 4.7E-6
Default unit: s
Usage: Asynchronous command
24.9.15 Trigger qualification
The A-event can have its own trigger qualification. Qualification is not available for B-
and R-events (event suf
fix m = 2 | 3) and some trigger types.
A suffix defines the trigger type to which the qualification belongs.
Table 24-3: Trigger type suffixes
Suffix Trigger type
1 EDGE
2 GLITch
3 WIDTh
4 RUNT
5 WINDow
6 TIMeout
7 INTerval
8 Qualification is not supported (SLEWrate)
9 Qualification is not supported (SETHold)
10 STATe
11 PATTern
12 qualification is not supported (ANEDge, ext. trigger input)
13 Currently not used
14 Currently not used
15 Qualification is not supported (SERPattern)
16 Qualification is not supported (NFC)
Trigger