Remote control commands
R&S
®
RTO6
1418User Manual 1801.6687.02 ─ 05
Suffix Trigger type
17 Qualification is not supported (TV)
18 Qualification is not supported (CDR)
TRIGger<m>:QUALify<n>:STATe...................................................................................1418
TRIGger<m>:
QUALify<n>:A[:ENABle]............................................................................1418
TRIGger<m>:QUALify<n>:B[:ENABle]............................................................................1418
TRIGger<m>:QUALify<n>:C[:ENABle]............................................................................1418
TRIGger<m>:QUALify<n>:D[:ENABle]............................................................................1418
TRIGger<m>:QUALify<n>:A:LOGic................................................................................1419
TRIGger<m>:QUALify<n>:B:LOGic................................................................................1419
TRIGger<m>:QUALify<n>:C:LOGic............................................................................... 1419
TRIGger<m>:QUALify<n>:D:LOGic............................................................................... 1419
TRIGger<m>:QUALify<n>:AB:LOGic..............................................................................1420
TRIGger<m>:QUALify<n>:CD:LOGic............................................................................. 1420
TRIGger<m>:QUALify<n>:ABCD:LOGic.........................................................................1420
TRIGger<m>:QUALify<n>:STATe <AddTrigLogi>
Enables the use of the qualification definition for the selected trigger event.
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
<n> 1..18
Defines the trigger type, see Table 24-3.
Parameters:
<AddT
rigLogi> ON | OFF
ON
The qualification expression is considered for the trigger event.
OFF
The qualification expression is ignored for the trigger event.
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:QUALify<n>:A[:ENABle] <State>
TRIGger<m>:QUALify<n>:B[:ENABle] <State>
TRIGger<m>:QUALify<n>:C[:ENABle] <State>
TRIGger<m>:QUALify<n>:D[:ENABle] <State>
The command is relevant for qualification, for pattern trigger and for state trigger.
Select the channels to be considered:
A[:ENABle]: C1
B[:ENABle]: C2
C[:ENABle]: C3