Remote control commands
R&S
®
RTO6
1419User Manual 1801.6687.02 ─ 05
D[:ENABle]: C4
The trigger source cannot be enabled.
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
<n> 1..18
Trigger type, see
Table 24-3.
Parameters:
<State> ON | OFF
ON
The qualification expression is considered.
OFF
The qualification expression is ignored.
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:QUALify<n>:A:LOGic <Operator>
TRIGger<m>:QUALify<n>:B:LOGic
<Operator>
TRIGger<m>:QUALify<n>:C:LOGic <Operator>
TRIGger<m>:QUALify<n>:D:LOGic <Operator>
Defines the logic for the indicated channel:
A: C1
B: C2
C: C3
D: C4
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
<n> 1..18
Trigger type, see
Table 24-3.
Parameters:
<Operator> DIRect | HIGH | NOT | LOW
DIRect
Input value remains unchanged
NOT
Input value is inverted
*RST: DIRect
Usage: Asynchronous command
Trigger