Remote control commands
R&S
®
RTO6
1640User Manual 1801.6687.02 ─ 05
24.16  Data management
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order
,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
T
o prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
●
www.rohde-schwarz.com/rc-via-scpi, chapter "Command Sequence and Synchro-
nization"
●
Chapter 24.3, "Programming examples", on page 1259
● Instrument settings..............................................................................................1640
● Savesets and one file..........................................................................................1648
● Waveform data transmission...............................................................................1649
● Waveform data export to file............................................................................... 1649
● Waveform histogram export to file...................................................................... 1656
● Eye......................................................................................................................1657
● Long term measurement results and measurement histogram export to file......1659
● Screenshots........................................................................................................ 1660
● Reports................................................................................................................1664
● Autonaming.........................................................................................................1666
● CSV export..........................................................................................................1668
● External application.............................................................................................1668
24.16.1  Instrument settings
The Mass MEMory subsystem provides commands to access the storage media and to
save and reload instrument settings.
File and directory names
The <file_name> and <directory_name> parameters are strings. Some commands use
a fixed directory; for others the <file_name> can contain the complete path including
the drive name and all subdirectories. If no complete path is specified, the file location
is relative to the current directory
, queried with MMEMory:
CDIRectory?. The file
name itself may contain the period as a separator for extensions.
Example: 'C:\TEMP\TRASH\test.txt' for the file named test.txt in the
TEMP\TRASH subdirectory of the internal hard disk drive C:\.
File and directory names can be chosen according to Windows™ conventions; the
restrictions placed on file names known from DOS systems do not apply. All letters and
numbers are allowed, as well as the special characters "_", "^", "$", "~", "!", "#", "%",
"&", "-", "{", "}", "(", ")", "@" and "`". Reserved file names are CON, AUX, COM1, ...,
COM4, LPT1, ..., LPT3, NUL and PRN.
The use of wildcards ? and * is not allowed.
Data management