EasyManua.ls Logo

R&S RTO6 - 24.16 Data management; 24.16.1 Instrument settings

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1640User Manual 1801.6687.02 ─ 05
24.16 Data management
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order
,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
T
o prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
www.rohde-schwarz.com/rc-via-scpi, chapter "Command Sequence and Synchro-
nization"
Chapter 24.3, "Programming examples", on page 1259
Instrument settings..............................................................................................1640
Savesets and one file..........................................................................................1648
Waveform data transmission...............................................................................1649
Waveform data export to file............................................................................... 1649
Waveform histogram export to file...................................................................... 1656
Eye......................................................................................................................1657
Long term measurement results and measurement histogram export to file......1659
Screenshots........................................................................................................ 1660
Reports................................................................................................................1664
Autonaming.........................................................................................................1666
CSV export..........................................................................................................1668
External application.............................................................................................1668
24.16.1 Instrument settings
The Mass MEMory subsystem provides commands to access the storage media and to
save and reload instrument settings.
File and directory names
The <file_name> and <directory_name> parameters are strings. Some commands use
a fixed directory; for others the <file_name> can contain the complete path including
the drive name and all subdirectories. If no complete path is specified, the file location
is relative to the current directory
, queried with MMEMory:
CDIRectory?. The file
name itself may contain the period as a separator for extensions.
Example: 'C:\TEMP\TRASH\test.txt' for the file named test.txt in the
TEMP\TRASH subdirectory of the internal hard disk drive C:\.
File and directory names can be chosen according to Windows™ conventions; the
restrictions placed on file names known from DOS systems do not apply. All letters and
numbers are allowed, as well as the special characters "_", "^", "$", "~", "!", "#", "%",
"&", "-", "{", "}", "(", ")", "@" and "`". Reserved file names are CON, AUX, COM1, ...,
COM4, LPT1, ..., LPT3, NUL and PRN.
The use of wildcards ? and * is not allowed.