Remote control commands
R&S
®
RTO6
2520User Manual 1801.6687.02 ─ 05
24.23.1 General
POW
er:
ENABle........................................................................................................... 2520
POWer:SOURce:CURRent<1..2>..................................................................................2520
POWer:SOURce:VOLTage<1..4>...................................................................................2520
POWer:ENABle
Activates the power mode and initializes the power measurements. If the power mode
is disabled, the instrument does not accept any POWer command.
Use POWer:ENABle after each *RST.
Example: See Chapter 24.3.8.1, "Auto deskew", on page 1277
Usage: Event
Asynchronous command
POWer:SOURce:CURRent<1..2> <CurrentSource>
Sets the channel for the current source.
Parameters:
<CurrentSource> CHAN1 | CHANnel1 | CHAN2 | CHANnel2 | CHAN3 |
CHANnel3 | CHAN4 | CHANnel4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
*RST
: CURRent1: CHAN2, CURRent2: CHAN4,
Usage: Asynchronous command
POWer:SOURce:VOLTage<1..4> <VoltageSource>
Sets the channel for the voltage source input.
Parameters:
<V
oltageSource> CHAN1 | CHANnel1 | CHAN2 | CHANnel2 | CHAN3 |
CHANnel3 | CHAN4 | CHANnel4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
*RST
: VOLTage1: CHAN1, VOLTage2: CHAN3, VOLT-
age3: CHAN3, VOLT
age4: CHAN4
Usage: Asynchronous command
24.23.2 Deskew
Programming example: Chapter 24.3.8.1, "Auto deskew", on page 1277
Power analysis (option R&S RT