19.4.1.2 Configuring modulation analysis
For details of the configuration settings, see Chapter 19.4.1.3, "Modulation analysis
settings", on page
1
154.
1. Select "Analysis" > "Power".
2. Under "Switching / Control Loop", select "Modulation Analysis".
3. Connect the differential voltage probe and the current probe to the oscilloscope.
4. Deskew the probes as described in Chapter 19.1.1.1, "Auto deskew",
on page 1123.
5. Connect the probes to the DUT as shown in the "Channels" tab:
6. Select the correct channels for the "Current Source" and the "Voltage Source".
7. Select the "Settings" tab.
8.
Set the "Source" and the "Type" of measurement.
9. Select an "Optimize Scaling" option.
Control loop