EasyManua.ls Logo

R&S RTO6 - 24.18.4.2 Edge trigger

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2433User Manual 1801.6687.02 ─ 05
Parameters:
<Type> EDGE | WIDTh | TIMeout | SETHold | STATe | PATTern |
SERPattern |
SERPattern = serial pattern trigger
*RST: EDGE
Usage: Asynchronous command
TRIGger<m>:PARallel:SETHold:CSOurce[:VALue] <ClockSource>
TRIGger<m>:PARallel:ST
ATe:CSOurce:VALue <ClockSource>
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue] <ClockSource>
Selects the digital channel of the clock signal.
Parameters:
<ClockSource> D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 |
D13 | D14 | D15
*RST: D0
Usage: Asynchronous command
TRIGger<m>:PARallel:EDGE:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:P
ARallel:WIDTh:
EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:TIMeout:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:STATe:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:PATTern:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:SPATtern:EXPRession[:DEFine] <LogicalExpr>
Defines a logical combination of several digital channels as trigger condition if
TRIGger<m>:SOURce[:SELect] is set to LOGIC.
Parameters:
<LogicalExpr> String with logical expression
Example:
TRIGger:PARallel:EDGE:EXPRession 'D1 and D2'
Usage: Asynchronous command
24.18.4.2 Edge trigger
See also:
TRIGger<m>:PARallel:
EDGE:EXPRession[:DEFine] on page 2433
TRIGger<m>:PARallel:EDGE:SLOPe.............................................................................2433
TRIGger<m>:PARallel:EDGE:SLOPe <Slope>
Defines the edge - the state transition - of the signal to trigger on a single digital chan-
nel (a logic bit), or a logical combination of digital channels.
Mixed signal option (MSO, R&S RT
O6-B1)

Table of Contents

Other manuals for R&S RTO6