TRIGger<m>:QUALify<n>:AB:LOGic <Operator>
TRIGger<m>:QUALify<n>:CD:LOGic <Operator>
TRIGger<m>:QUALify<n>:ABCD:LOGic <Operator>
Defines the logical combination of the indicated channels after evaluating the previous
logical operations:
AB: CH1 and CH2
CD: CH3 and CH4
ABCD: result of AB and CD
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n> 1..18
Trigger type, see
Table 24-3.
Parameters:
<Operator> AND | NAND | OR | NOR
AND
logical AND, conjunctive combination
NAND
logical NOT AND
OR
logical OR, disjunctive combination
NOR
logical NOT OR
*RST: AND
Usage: Asynchronous command
24.9.16 Holdoff
TRIGger<m>:HOLDof
f:MODE....................................................................................... 1420
TRIGger<m>:HOLDoff:TIME......................................................................................... 1421
TRIGger<m>:HOLDoff:EVENts......................................................................................1422
TRIGger<m>:HOLDoff:MIN...........................................................................................1422
TRIGger<m>:HOLDoff:MAX..........................................................................................1423
TRIGger<m>:HOLDoff:AUTotime?................................................................................. 1423
TRIGger<m>:HOLDoff:SCALing.................................................................................... 1423
TRIGger<m>:HOLDoff:MODE <Mode>
Selects the method to define the holdoff condition.
The trigger holdoff defines when the next trigger after the current will be recognized.
Thus, it affects the next trigger to occur after the current one. Holdoff helps to obtain
stable triggering when the oscilloscope is triggering on undesired events.