Remote control commands
R&S
®
RTO6
1422User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
TRIGger<m>:HOLDoff:EVENts <Events>
Defines the number of triggers to be skipped. The next trigger only occurs when this
number of events is reached. The setting is relevant if the holdof
f mode is set to
EVENts.
See also:
●
TRIGger<m>:
HOLDoff:MODE
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Events> Range:
1 to 2147483647
Increment: 1
*RST: 1
Example:
TRIGger:HOLDoff:MODE EVENts
TRIGger:
HOLDoff:
EVENts 5
Usage: Asynchronous command
TRIGger<m>:HOLDoff:MIN <RandomMinTime>
Defines the lower limit for the random time holdof
f. The setting is relevant if the holdoff
mode is set to RANDom.
See also:
●
TRIGger<m>:HOLDoff:MODE
●
TRIGger<m>:HOLDoff:MAX
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
Parameters:
<RandomMinT
ime> Range: 100E-9 to 5
Increment: 200E-6
*RST: 1E-3
Default unit: s
Example:
TRIGger:HOLDoff:MODE RANDom
TRIGger:
HOLDoff:MIN 1ms
TRIGger:HOLDoff:MAX 2ms
The holdoff time is set randomly between 1 ms and 2 ms.