Compliance tests
R&S
®
RTO6
1218User Manual 1801.6687.02 ─ 05
●
100BASE-T1: (R&S RT
O6-K24)
●
1000BASE-T1: (R&S RTO6-K87)
●
1000BASE-T1: (R&S RTO6-K88
●
10BASE-T1: (R&S RTO6-K89)
The test fixture and the Ethernet compliance test procedures are described in the sep-
arate manuals:
●
R&S®RTO, R&S®RTP Ethernet Compliance Test, User Manual"
●
"R&S®RT-ZF2 Ethernet Compliance Test Fixture Set Manual"
●
"R&S®RT-ZF3 Frequency Converter Manual
●
"R&S®RT-ZF4 10BASE-Te Fixture Manual
●
R&S®RT-ZF5 Ethernet Probing Fixture Manual
●
R&S®RT-ZF6 Frequency Converter Manual
21.2.1 Test equipment
For Ethernet compliance tests, the following test equipment is needed:
●
The free-of-charge R&S ScopeSuite software, which can be installed on a com-
puter or directly on the oscilloscope.
●
10/100/1000BASE-T Ethernet compliance tests:
– R&S RTO/R&S RTP oscilloscope with at least 600 MHz bandwidth
– Differential probe with at least 1 GHz bandwidth
– R&S RTO/R&S RTP-K22 10/100/1000BASE-T Ethernet compliance test option
(required option, installed on the oscilloscope)
– R&S RT-ZF2 Ethernet test fixture set
– For energy-efficient Ethernet tests, in addition:
R&S RT-ZF5 Ethernet test fixture set for 100/1000BASE-T EEE tests
R&S RT-ZF4 test fixture for 10BASE-Te tests
●
2.5GBASE-T Ethernet compliance tests:
– R&S RTO/R&S RTP oscilloscope with at least 600 MHz bandwidth
For precise power spectral density and power level measurements up to
1 GHz, it is recommended to use an oscilloscope and a differential probe with
1 GHz bandwidth each.
– Differential probe with at least 1GHz bandwidth
– R&S RTO/R&S RTP-K23 GBASE-T Ethernet compliance test option (required
option, installed on the R&S RTO6)
– R&S RT-ZF2 Ethernet test fixture set
●
5GBASE-T Ethernet compliance tests:
– R&S RTO/R&S RTP oscilloscope with at least 1 GHz bandwidth
For precise power spectral density and power level measurements up to
1.5 GHz, it is recommended to use an oscilloscope and a differential probe with
1.5 GHz bandwidth each.
– Differential probe with at least 1.5 GHz bandwidth
Ethernet compliance tests (options R&S R
T