EasyManua.ls Logo

R&S RTO6 - 24.8.7.8 Current probes

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1362User Manual 1801.6687.02 ─ 05
FREE
Any other probe that is not recognized by the instrument.
*RST: FREE
Usage: Asynchronous command
PROBe<m>:SETup:OFFSet:TOMean
Performs an automatic compensation for a DC component of the specified input signal
using the result of a background mean measurement.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
24.8.7.8 Current probes
To set up R&S
RT-ZC10 and R&S RT-ZC20, use PROBe<m>:SETup:ATTenuation:
DEFProbe.
PROBe<m>:SETup:DEGauss....................................................................................... 1362
PROBe<m>:SETup:OFFSet:STPRobe...........................................................................1362
PROBe<m>:SETup:OFFSet:ZADJust.............................................................................1363
PROBe<m>:SETup:DEGauss
Demagnetizes the core if it has been magnetized by switching the power on and off, or
by an excessive input. Always carry out demagnetizing before measurement. The
demagnetizing process takes about one second.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
Firmware/software: FW 2.50
PROBe<m>:SETup:OFFSet:STPRobe
Saves the zero adjust value in the probe box. If you connect the probe to another
channel or to another R&S RTx oscilloscope, the value is read out again.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command