Triggers
R&S
®
RTO6
234User Manual 1801.6687.02 ─ 05
Use existing masks for zone trigger
Prerequisites: At least one mask is created and visible on the display
.
1.
Tap the "Zone trigger" icon on the toolbar.
2. In the signal bar, select if the signal must violate the zone to cause a trigger (Must
Intersect), or vice versa (Must Not Intersect).
3. Tap the mask on the display.
4. Tap "Use selected mask" in the signal bar.
"MT1" is shown in the expression field.
5. If you want to add another mask test or zone, select the logic combination of the
zones AND (both zones) or OR (either zone).
6. To add another mask, tap the mask on the display.
Tap "Use selected mask" in the signal bar.
7. To add another zone, tap the corner points of the zone on the screen.
If all points are set, tap "Finish zone" in the signal bar.
If a zone trigger is already defined, you can also add a new mask to the zone trigger
when creating the mask.
Intersect
Defines if the signal must intersect the zone to allow the instrument to trigger, or if it
must not intersect the zone.
Combine
Sets the logic combination of two zones.
Use selected mask
Includes the selected zone or mask in the zone expression.
Clear Expr
Deletes the zone trigger expression. The zones are not deleted, they remain as usual
masks, and the result boxes of the mask test appear.
Delete all
Deletes the zone trigger expression and the zones.
6.9.3 Advanced setup and analysis with zone trigger
Access: overlay menu for zone trigger > "Advanced Setup", or "Trigger" menu >
"Setup" > select sequence with zone trigger > "Setup Z trigger" dialog
Zone trigger