Acquisition and waveform setup
R&S
®
RTO6
183User Manual 1801.6687.02 ─ 05
In the diagram, you see the output waveforms, and the signal icons of the input chan-
nels are grayed out. The input channels are only visible if
"P"
or "N" are selected as
output.
"P", "N"
Displays the waveform of the positive or negative input signal.
Without deembedding, the input channel is shown directly, no addi-
tional waveform is created. If dif
ferential deembedding is active, you
always see the deembedded waveforms, but never the captured input
signal.
"Differential"
Displays the differential signal as a new waveform.
"Common"
Displays the common mode signal as a new waveform.
Remote command:
DIFFerential<m>:
AOUTput
on page 1380
DIFFerential<m>:BOUTput on page 1380
If deembedding by software is active, use the following commands:
DIFFerential<m>:COMMon:STATe on page 1381
DIFFerential<m>:DIFFerential:STATe on page 1381
DIFFerential<m>:NSIGnal:STATe on page 1381
DIFFerential<m>:PSIGnal:STATe on page 1381
5.8.2 Vertical setup of differential signals
Access: "Menu" > "Vertical" > "Differential" > "Vertical Setup"
Scale
Sets the vertical scale of differential and common mode waveforms.
V
ertical settings of P and N output are the vertical settings of the input channels.
Remote command:
DIFFerential<m>:COMMon:SCALe on page 1381
DIFFerential<m>:DIFFerential:SCALe on page 1381
Differential signals