Acquisition and waveform setup
R&S
®
RTO6
182User Manual 1801.6687.02 ─ 05
Figure 5-19: Setup of deembedded differential signals
Enable
Activates the differential signal.
Remote command:
DIFFerential<m>:
STATe on page 1380
P, N
Select the analog input channel of the positive and negative signal.
Remote command:
DIFFerential<m>:PSIGnal[:SELect] on page 1380
DIFFerential<m>:NSIGnal[:SELect] on page 1380
Output: P, N, Diff, Common
Select the waveforms that are displayed as result of differential processing.
The number of output waveforms per differential signal depends on the deembedding:
●
No deembedding: 2 output waveforms.
●
With deembedding: up to 4 output waveforms. All output waveforms can be dis-
played at the same time.
Differential signals