Remote control commands
R&S
®
RTO6
1902User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<CRCCalculation> SAEJ | TLE
SAEJ: according to the standard
TLE: according to the computing method for TLE_4998X sen-
sors.
*RST: SAEJ
BUS<m>:SENT:PPULse <PausePulse>
Determines if a pause pulse is part of the SENT transmission sequence.
Suffix:
<m>
.
1..4
Selects the serial data bus.
Parameters:
<PausePulse> NPP | PP | PPFL
NPP
T
ransmits the SENT message without pause pulse.
PP
T
ransmits the message with a fixed pulse length, automatically
calculated.
PPFL
Transmits the pause pulse with a user-defined frame length to
obtain a transmission sequence with constant length.
*RST: PPFL
BUS<m>:SENT:PPFLength <FrameLength>
Defines a constant transmission sequence length. To select the fixed sequence length,
set
BUS:SENT:PPUL PPFL.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<FrameLength> Range: 104 to 922
Increment: 1
*RST: 256
24.17.11.2 Trigger
Event in a trigger sequence: 1 = A-event only
Protocols