Remote control commands
R&S
®
RTO6
1903User Manual 1801.6687.02 ─ 05
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
T
o trigger on a serial bus, make sure that:
●
TRIGger<m>:
SOURce[:SELect] is set to SBUS.
●
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
●
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:SENT:TYPE.............................................................................................1903
TRIGger<m>:SENT:TTYPe........................................................................................... 1904
TRIGger<m>:SENT:STATus.......................................................................................... 1904
TRIGger<m>:SENT:TDCN............................................................................................ 1905
TRIGger<m>:SENT:TDMN............................................................................................1905
TRIGger<m>:SENT:TDMX............................................................................................1905
TRIGger<m>:SENT:STYPe...........................................................................................1905
TRIGger<m>:SENT:SIDType.........................................................................................1906
TRIGger<m>:SENT:SICN..............................................................................................1906
TRIGger<m>:SENT:SIMN............................................................................................. 1906
TRIGger<m>:SENT:SIMX............................................................................................. 1906
TRIGger<m>:SENT:SDCN............................................................................................1907
TRIGger<m>:SENT:SDMN............................................................................................1907
TRIGger<m>:SENT:SDMX............................................................................................1907
TRIGger<m>:SENT:FORMerror.....................................................................................1907
TRIGger<m>:SENT:PULSeerror.................................................................................... 1908
TRIGger<m>:SENT:PPERioderror................................................................................. 1908
TRIGger<m>:SENT:CRCerror....................................................................................... 1908
TRIGger<m>:SENT:IRFLength...................................................................................... 1909
TRIGger<m>:SENT:TYPE <Type>
Selects the trigger event for the SENT transmission type.
Parameters:
<Type> CALI | TSEQ | SMSG | ERRC
CALI
CALIbration: triggers on the falling edge of the calibration/
synchronization pulse.
TSEQ
T
ransmission SEQuence: triggers either on the falling edge of
the status nibble, or on the last data nibble.
T
o set the transmission sequence conditions, use
TRIGger<m>:SENT:TTYPe and TRIGger<m>:SENT:STATus.
To set the data condition, use TRIGger<m>:SENT:TDCN,
BUS<m>:SENT:DNIBbles, TRIGger<m>:SENT:TDMN and
TRIGger<m>:SENT:TDMX.
Protocols