Remote control commands
R&S
®
RTO6
1906User Manual 1801.6687.02 ─ 05
IDDT
T
riggers at the end of the combination of identifier and data.
T
o set the identifier condition, use the commands shown above.
To set the data condition, use TRIGger<m>:SENT:SDCN,
TRIGger<m>:SENT:SDMN and TRIGger<m>:SENT:SDMX.
*RST: ID
TRIGger<m>:SENT:SIDType <SMIDType>
Sets the message ID format (4 bit or 8 bit) of the enhanced serial message.
Parameters:
<SMIDT
ype> B4 | B8
*RST: B4
TRIGger<m>:SENT:SICN <SMIDOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<SMIDOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one data pattern
to be set with TRIGger<m>:SENT:SIMN.
INRange | OORange
In range / Out of range. To define the range set the minimum
and maximum values with TRIGger<m>:SENT:SIMN and
TRIGger<m>:SENT:SIMX.
*RST: EQUal
TRIGger<m>:SENT:SIMN <SMIDPattern>
Specifies a message identifier pattern, or sets the start value of an identifier range.
Parameters:
<SMIDPattern> Numeric or string pattern, Chapter 24.4.6, "Bit pattern parame-
ter", on page 1286. The string parameter accepts the bit value X
(don't care).
TRIGger<m>:SENT:SIMX <SMIDPatternTo>
Sets the end value of an identifier range for TRIGger<m>:SENT:SICN INRange or
OORange.
Protocols