EasyManua.ls Logo

R&S RTO6 - 24.8.11 Skew

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1384User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
TRIGger<m>:RFReject <Bandwidth>
Suffix:
<m>
.
1..3
Irrelevant, omit the suf
fix.
Parameters:
<Bandwidth>
Range: 100E+3 to 2E+9
Increment: 1000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
24.8.11 Skew
CHANnel<m>:SKEW:MANual.......................................................................................
1384
CHANnel<m>:SKEW:TIME........................................................................................... 1384
PROBe<m>:SKEState..................................................................................................1385
CHANnel<m>:SKEW:EXTended.................................................................................... 1385
CHANnel<m>:SKEW:MANual <ManualCompens>
If enabled, the skew offset value (CHANnel<m>:SKEW:TIME) is used for compensa-
tion. This improves horizontal and trigger accuracy.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualCompens> ON | OFF
*RST: ON
Usage: Asynchronous command
CHANnel<m>:SKEW:TIME <Offset>
Sets an delay value, that is known from the circuit specifics but cannot be compensa-
ted by the instrument automatically
. It affects only the selected input channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Acquisition and setup

Table of Contents

Other manuals for R&S RTO6

Related product manuals