Remote control commands
R&S
®
RTO6
2040User Manual 1801.6687.02 ─ 05
Return values:
<FrameByteValue> Range: 0 to 255
Increment: 1
*RST: 0
Usage: Query only
24.17.17 MDIO (option R&S RTO6-K540)
Configuration.......................................................................................................2040
T
rigger.................................................................................................................2042
Decode results.................................................................................................... 2045
Search settings................................................................................................... 2049
Search results..................................................................................................... 2051
24.17.17.1 Configuration
In all BUS<m>:MDIO commands, the suffix <m> selects the serial bus.
BUS<m>:MDIO:CLOCk:SOURce...................................................................................2040
BUS<m>:MDIO:DATA:SOURce.....................................................................................2041
BUS<m>:MDIO:CLOCk:THReshold:HIGH...................................................................... 2041
BUS<m>:MDIO:CLOCk:THReshold:LOW.......................................................................2041
BUS<m>:MDIO:DATA:THReshold:HIGH.........................................................................2041
BUS<m>:MDIO:DATA:THReshold:LOW......................................................................... 2042
BUS<m>:MDIO:PRESet............................................................................................... 2042
BUS<m>:MDIO:COUPling............................................................................................ 2042
BUS<m>:MDIO:CLOCk:SOURce <SourceClock>
Selects the source for the clock line (management data clock, MDC). Permitted selec-
tions are the analog channels "C1""C4" and the digital channels "D0""D15".
Suffix:
<m>
.
1..4
Parameters:
<SourceClock> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 |
R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data and clock lines. For triggering on a serial bus, analog or
digital input channels are required.
See Chapter 24.4.2, "Waveform parameter", on page 1284
*RST: C2W1