Remote control commands
R&S
®
RTO6
2466User Manual 1801.6687.02 ─ 05
24.20.2 TDR/TDT analysis
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order
,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
T
o prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
www.rohde-schwarz.com/rc-via-scpi, chapter "Command Sequence and Synchro-
nization"
Chapter 24.3, "Programming examples", on page 1259
24.20.2.1 Calibration
TDRT:DEFault............................................................................................................. 2466
TDRT:CALibration:OPEN:NAME....................................................................................2466
TDRT:CALibration:OPEN[:EXECute].............................................................................. 2466
TDRT:TYPE.................................................................................................................2467
TDRT:PULSe:OUT....................................................................................................... 2467
TDRT:PULSe:REFerence..............................................................................................2467
TDRT:SLENgth............................................................................................................ 2467
TDRT:CALibration:SAVE:NAME.....................................................................................2468
TDRT:CALibration:SAVE[:EXECute]...............................................................................2468
TDRT:DEFault
Starts the measurement with default correction values instead of calibrating the setup.
Default values are rough values without time reference.
Usage: Event
Asynchronous command
TDRT:CALibration:OPEN:NAME <Path>
Sets the file name, file format and path of the TDR/TDT calibration file to be opened.
The file format is always .xml.
Parameters:
<Path> String with path and file name with extension .xml
Usage: Asynchronous command
TDRT:CALibration:OPEN[:EXECute]
Opens and loads the TDR/TDT calibration from the file specified with TDRT:
CALibration:OPEN:
NAME.
Pulse Source and TDR/TDT Analysis (options R&S RT
O6-B7/K130)