EasyManua.ls Logo

R&S RTO6 - 24.19.1.2 Function generator

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2444User Manual 1801.6687.02 ─ 05
Usage: Event
24.19.1.2 Function generator
WGENerator<m>:FUNCtion[:
SELect]
.............................................................................2444
WGENerator<m>:FREQuency.......................................................................................2444
WGENerator<m>:PERiod............................................................................................. 2444
WGENerator<m>:FUNCtion:PULSe[:WIDTh].................................................................. 2445
WGENerator<m>:FUNCtion:RAMP[:SYMMetry]..............................................................2445
WGENerator<m>:FUNCtion:SQUare:DCYCle.................................................................2445
WGENerator<m>:FUNCtion[:SELect] <FunctionType>
Sets the type of waveform to be generated for the function generator.
Suffix:
<m>
.
1..2
Parameters:
<FunctionType> SINusoid | SQUare | RAMP | DC | PULSe | SINC | CARDiac |
GAUSs |
LORNtz | EXPRise | EXPFall
SINC: Cardial sine
*RST: SINusoid
WGENerator<m>:FREQuency <Frequency>
Sets the frequency of the waveform. The available frequency range depends on the
selected function type.
Suffix:
<m>
.
1..2
Parameters:
<Frequency> Range: 1E-3 to 100E+6
Increment: 1
*RST: 1E+6
Default unit: Hz
WGENerator<m>:PERiod <Period>
Sets the period of the waveform. The available period range depends on the selected
function type.
Suffix:
<m>
.
1..2
Parameters:
<Period> Range: 8E-9 to 1000
Increment: 1
*RST: 1E-6
Default unit: s
Waveform generator (option R&S RT