Triggers
R&S
®
RTO6
201User Manual 1801.6687.02 ─ 05
"Within"
Triggers if the runt length is inside a given time range. The range is
defined by 
"Runt width" and "
±Delta".
"Outside"
Triggers if the runt length is outside a given time range. The range
definition is the same as for "Within" range.
Remote command: 
TRIGger<m>:RUNT:
RANGe on page 1396
Runt width
For the ranges "Shorter" and "Longer", the runt width defines the maximum and mini-
mum pulse width, respectively.
For the ranges "Within" and "Outside", the runt width defines the center of a range
which is defined by "±Delta".
Remote command: 
TRIGger<m>:RUNT:WIDTh on page 1397
±Delta
Defines a range around the given runt width.
Remote command: 
TRIGger<m>:RUNT:DELTa on page 1397
6.3.6  Window
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Window"
The window trigger checks the signal run in relation to a "window". The window is
formed by the upper and lower voltage levels. The trigger condition is fulfilled, if the
waveform enters or leaves the window, or if the waveform stays inside or outside for a
time longer or shorter than specified.
With the window trigger
, you can display longer transient effects.
Trigger types