Triggers
R&S
®
RTO6
202User Manual 1801.6687.02 ─ 05
Vertical condition
Selects how the signal run is compared with the window:
"Enter"
Triggers when the signal crosses the upper or lower level and thus
enters the window made up of these two levels.
"Exit"
Triggers when the signal leaves the window.
"Stay within"
Triggers if the signal stays between the upper and lower level for a
specified time. The time is defined in various ways by the Time condi-
tion
.
"Stay outside"
Triggers if the signal stays above the upper level or below the lower
level for a specified time. The time is also defined by the "T
ime condi-
tion".
Remote command: 
TRIGger<m>:WINDow:
RANGe on page 1398
Upper level
Sets the upper voltage limit for the window.
Remote command: 
TRIGger<m>:LEVel<n>:WINDow:UPPer on page 1398
Lower level
Sets the lower voltage limit for the window.
Remote command: 
TRIGger<m>:LEVel<n>:WINDow:LOWer on page 1398
Time condition
Selects how the time limit of the window is defined. Time conditioning is available for
the vertical conditions "Stay within" and "Stay outside".
"Within"
Triggers if the signal stays inside or outside the vertical window limits
at least for the time Width - Delta
 and for Width + Delta at the most.
"Outside"
"Outside" is the opposite definition of "Within". The instrument triggers
if the signal stays inside or outside the vertical window limits for a
time shorter than Width - Delta or longer than 
Width + Delta.
"Shorter"
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
"Longer"
Triggers if the signal crosses vertical limits after the specified "Width"
time is reached.
Remote command: 
TRIGger<m>:
WINDow:
TIME on page 1399
Width
For the ranges "Within" and "Outside", the width defines the center of a time range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time
lapse, respectively.
Trigger types