Remote control commands
R&S
®
RTO6
1399User Manual 1801.6687.02 ─ 05
ENTer
T
riggers when the signal crosses the upper or lower level and
thus enters the window made up of these two levels.
EXIT
T
riggers when the signal leaves the window.
WITHin
Triggers if the signal stays between the upper and lower level for
a specified time. The time is defined using the TRIGger<m>:
WINDow:TIME command.
OUTSide
Triggers if the signal stays above the upper level or below the
lower level for a specified time. The time is defined using the
TRIGger<m>:WINDow:TIME command.
*RST: ENTer
Usage: Asynchronous command
TRIGger<m>:WINDow:TIME <T
imeRangeMode>
Defines the limit of the window in relation to the time specified using TRIGger<m>:
WINDow:
WIDTh and TRIGger<m>:WINDow:DELTa. Time conditioning is available for
TRIGger<m>:WINDow:RANGe= "WITHin" and "OUTSide".
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<TimeRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers if the signal stays inside or outside the vertical window
limits at least for the time
Width - Delta and for Width + Delta at
the most.
OUTSide
"Outside" is the opposite definition of "Within". The instrument
triggers if the signal stays inside or outside the vertical window
limits for a time shorter than Width - Delta or longer than Width +
Delta.
SHORter
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
*RST: WITHin
Usage: Asynchronous command
Trigger