R&S RTO6 - 24.18.3 Digital resolution

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2430User Manual 1801.6687.02 ─ 05
BUS<m>:PARallel:DECTable<n>:SHOW <ShowDecodeT
able>
A
vailable only if BUS<m>:PARallel:CLON is set to ON.
If enabled, a results box opens with decoded values of the bus signal and its time. You
can query the number of values (BUS<m>:PARallel:DECTable<n>:COUNt?) and
their data value (BUS<m>:PARallel:DECTable<n>:DATA?).
Suffix:
<m>
.
1..4
<n> 1..5
Parameters:
<ShowDecodeTable> ON | OFF
*RST: OFF
BUS<m>:PARallel:DECTable<n>:COUNt?
Returns the number of clock edges in the decode table.
Suffix:
<m>
.
1..4
<n> 1..5
Usage: Query only
BUS<m>:PARallel:DECTable<n>:DATA?
Returns the data of the decoded values of the bus signal and its time.
Suffix:
<m>
.
1..4
<n> 1..5
Usage: Query only
24.18.3 Digital resolution
ACQuire:DRESolution?.................................................................................................
2430
ACQuire:POINts:DVALue?............................................................................................2431
ACQuire:DRESolution?
Returns the current digital resolution of the digital channels.
Return values:
<DigRes> Default unit: s
Usage: Query only
Asynchronous command