Waveform analysis
R&S
®
RTO6
274User Manual 1801.6687.02 ─ 05
●
It sets the number of acquisitions to be acquired in a fast segmentation acquisition
series. Thus, you can acquire exactly one fast segmentation acquisition series with
[Single].
If fast segmentation is enabled and configured to acquire the maximum number of
acquisitions, the acquisition count is set to that maximum number and cannot be
changed.
See also "Number of acquisitions
"
on page 132.
●
It is the "Finished" criteria for the state of a mask test.
Remote command:
ACQuire:COUNt on page 1334
Reset
Forces the immediate restart of the envelope and average calculation for all wave-
forms.
Remote command:
ACQuire:ARESet:IMMediate on page 1335
Arithmetic Reset > Mode
Defines when the envelope and average evaluation restarts.
"None"
No restart, the number of acquisitions considered by the waveform
arithmetics is not limited.
"Time"
Restarts the envelope and average calculation after the time defined
in "T
ime".
"W
aveforms"
Restarts the envelope and average calculation after a number of
acquired waveforms defined in "Count".
Remote command:
ACQuire:ARESet:MODE
on page 1335
ACQuire:ARESet:TIME on page 1335
ACQuire:ARESet:COUNt on page 1335
7.4 History
The history accesses the data of previous acquisitions and provides them for further
analysis.
7.4.1 About history
If a continuous acquisition runs, the captured data is stored in the sample memory and
the current acquisition is processed and shown on the display
. After the acquisition
was stopped, the history accesses the captured samples that were stored, displays
these samples as history waveforms, and makes them available for further analysis. It
considers all channels that were enabled during the running acquisition. When a new
acquisition is started with
[Run Stop] or [Single], the memory is cleared and written
anew.
History