Remote control commands
R&S
®
RTO6
1334User Manual 1801.6687.02 ─ 05
PDETect
Peak Detect: the minimum and the maximum of n samples in a
sample interval are recorded as waveform points.
HRESolution
High resolution: The average of n sample points is recorded as
waveform point.
RMS
The waveform point is the root mean square of n sample values.
*RST: SAMPle
Usage: Asynchronous command
CHANnel<m>[:WAVeform<n>]:ARIThmetics <Arithmetics>
Selects the method to build the resulting waveform from several consecutive acquisi-
tions of the signal. To define the number of acquisitions, use ACQuire:
COUNt.
See also: "Arithmetic" on page 130.
Suffix:
<m>
.
1..4
Selects the input channel.
<n> 1..3
Selects the waveform. If [:WA
Veform<n>] is omitted, waveform 1
is adressed.
Parameters:
<Arithmetics> OFF | ENVelope | AVERage
OFF
The data of the current acquisition is recorded according to the
decimation settings.
ENVelope
Detects the minimum and maximum values in an sample interval
over a number of acquisitions. T
o define the reset method,
use ...
AVERage
Calculates the average from the data of the current acquisition
and a number of acquisitions before.
*RST: OFF
Usage: Asynchronous command
ACQuire:COUNt <MaxAcqCnt>
The acquisition and average count has a double effect:
●
It sets the number of waveforms acquired with RUNSingle.
●
It defines the number of waveforms used to calculate the average waveform.
Acquisition and setup