Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1197User Manual 1801.6687.02 ─ 05
9. Tap "Execute".
10.
Set the positions of the cursors according to the measured spectrum.
On the screen, you can see the measurement of the spectrum. Also, the result box
shows the position of the peaks. For details, see Chapter 19.6.3.1, "Output spec-
trum results"
, on page 1195.
19.6.3.3 Output spectrum settings
In the "Channels" tab, you set the voltage source, see also: Chapter 19.2.1, "Channels
tab", on page 1134.
In the "Settings" tab, you configure the spectrum measurement parameters and display
settings.
Switching frequency
Sets the SMPS switching frequency. Set the value according to your DUT
.
Remote command:
POWer:SPECtrum:FREQuency on page 2557
Derived horizontal scale
Shows the scale of the displayed results. This value is calculated as the inverse of the
"SMPS switching frequency"