R&S
®
RT
O6
1206User Manual 1801.6687.02 ─ 05
For details, see:
"User skew" on page 124
"Skew of
fset" on page 124
20.2.6 Cable, adapter, fixture and custom settings
If you have added a cable, adapter, test fixture or other component to the measure-
ment setup, you set the characteristics in the corresponding subtabs of the "Deembed-
ding" > "Components" tab.
The characteristics are the same for these component types. Configuration differs
between 2-port and 4-port components, because 4-ports need additional settings. Gen-
erally, these components are configured by loading S-parameter files and analyzing its
characteristics.
To configure a component in a single-ended 2-port setup
1. Tap "Open".
2. Select and load the *.s2p Touchstone file that contains the S-parameters of the
component. The filetype is always *.s2p.
3. Assign the ports to the input and output of the component according to the data in
the Touchstone file. See also: "Input 1, Input 2, Output 1, Output 2" on page 1208.
4. Analyze the response curves of each S-parameter. The response curves are func-
tions of magnitude, phase, and group delay over frequency.
a) Select the S-parameter.
Deembedding settings