R&S
®
RTO6
1408User Manual 1801.6687.02 ─ 05
TRIGger<m>:LEVel<n>[:
VALue] on page 1391
TRIGger<m>:SCOupling on page 1407
To define the state, use the TRIG:QUAL...[ENABle] and TRIG:QUAL...LOGic
commands, which are described in Chapter 24.9.15, "Trigger qualification",
on page 1417.
24.9.12 Pattern trigger
The pattern trigger is only available for the A-event (Suffix = 1).
The pattern is defined using the commands:
TRIGger<m>:QUALify<n>:A[:ENABle] on page 1418
TRIGger<m>:QUALify<n>:A:LOGic on page 1419
TRIGger<m>:QUALify<n>:AB:LOGic on page 1420
These are the commands for channel 1, use the similar commands for channels 2, 3,
and 4.
TRIGger<m>:PATTern:MODE........................................................................................1408
TRIGger<m>:PATTern:TIMeout:MODE...........................................................................1409
TRIGger<m>:PATTern:TIMeout[:TIME]........................................................................... 1409
TRIGger<m>:PATTern:WIDTh:RANGe........................................................................... 1409
TRIGger<m>:PATTern:WIDTh[:WIDTh]...........................................................................1410
TRIGger<m>:PATTern:WIDTh:DELTa.............................................................................1410
TRIGger<m>:PATTern:MODE <Mode>
Adds additional time limitation to the pattern definition.
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
Parameters:
<Mode> OFF |
TIMeout | WIDTh
OFF
No time limitation. The event occurs if the pattern condition is
fulfilled.
TIMeout
Defines how long the result of the pattern condition must be true
or false. The duration of the timeout is defined using
TRIGger<m>:
PATTern:
TIMeout[:TIME].
WIDTh
Defines a time range for keeping up the true result of the pattern
condition. The range is defined using TRIGger<m>:PATTern:
WIDTh:RANGe.
*RST: OFF
Usage: Asynchronous command
Trigger