EasyManua.ls Logo

R&S RTO6 - 24.18.4.6 State trigger; 24.18.4.7 Pattern trigger

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2437User Manual 1801.6687.02 ─ 05
TRIGger<m>:PARallel:SETHold:STIMe <SetupT
ime>
Sets the minimum time
before the clock edge while data should be stable and not
change its state.
See also: "Setup time" on page 1003
Parameters:
<SetupTime> Range: -99.8E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
24.18.4.6 State trigger
See also:
TRIGger<m>:PARallel:
STATe:CSOurce:VALue on page 2433
TRIGger<m>:PARallel:STATe:EXPRession[:DEFine] on page 2433
TRIGger<m>:PARallel:STATe:CSOurce:EDGE................................................................2437
TRIGger<m>:PARallel:STATe:BIT<0..15>....................................................................... 2437
TRIGger<m>:PARallel:STATe:CSOurce:EDGE <Slope>
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time at which the logical states and the bus value are analyzed.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:PARallel:STATe:BIT<0..15> <Bit>
Sets the required state for each digital channel that is used in the bus.
Parameters:
<Bit> HIGH | LOW | DONTCARE | DONTcare
Bit value: 1 (HIGH), 0 (LOW), or X (DONTCARE = DONTcare)
Usage: Asynchronous command
24.18.4.7 Pattern trigger
TRIGger<m>:PARallel:PATTern:BIT<0..15> <Bit>
Sets the required state for each digital channel that is used in the bus.
Mixed signal option (MSO, R&S RT