Protocol analysis
R&S
®
RTO6
499User Manual 1801.6687.02 ─ 05
# ---------------------------------------------------------------------------
# Labels for I2C protocol
# Column order: Identifier type, Identifier value, Label
# ---------------------------------------------------------------------------
7,0x1E,Voltage
7,38h,Pressure
7,2Ah,Temperature
7,16h,Speed
7,118,Acceleration
7,07h,HighSpeed_Master_0x3
7,51h,EEPROM
10,3A2h,DeviceSetup
10,1A3h,GatewayStatus
10,06Eh,LeftSensor
# ----------------------------------------------------------------------------
13.2.2.3 Configuring I²C protocol
The configuration of the I²C is simple - assign the two lines to input channels, and set
the thresholds.
For details on configuration settings, see Chapter 13.2.2, "I²C configuration",
on page 496.
1.
Press the [Bus] key on the front panel.
2. Select the tab of the bus you want to set up, for example "SB1".
3. Select the "Setup" tab.
4. Tap the "Protocol" button and select the protocol: "I2C".
5. Tap the "SDA" button, and select the waveform of the data line.
6. Tap the "SCL" button, and select the waveform of the clock line.
7. Set the logical thresholds.
8. Enable "Decode".
13.2.3 I²C trigger
13.2.3.1 I²C trigger settings
Access: [Bus] > "Shortcuts" tab > "Setup trigger"
I²C (option R&S R
T
O6-K510)