EasyManua.ls Logo

R&S RTO6 - 24.18.4.8 Serial pattern trigger; 24.18.5 MSO data

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2440User Manual 1801.6687.02 ─ 05
TRIGger<m>:PARallel:PATTern:WIDTh:DELTa <WidthDelta>
Defines a range around the width value set with TRIGger<m>:
PARallel:
PATTern:
WIDTh[:WIDTh].
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 500E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
24.18.4.8 Serial pattern trigger
See also:
TRIGger<m>:
PARallel:
SPATtern:CSOurce[:VALue] on page 2433
TRIGger<m>:PARallel:SPATtern:EXPRession[:DEFine] on page 2433
TRIGger<m>:PARallel:SPATtern:CSOurce:EDGE <ClockEdge>
Sets the edge of the clock signal. The bit value is determined at the crossing of the
clock edge and the logical threshold.
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:PARallel:SPATtern:PATTern <Pattern>
Defines the serial bit string on which to trigger.
Parameters:
<Pattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The bit value X (don't care) is not
allowed.
Usage: Asynchronous command
24.18.5 MSO data
To export data of digital channels and parallel buses to file, use the following com-
mands:
EXPort:WAVeform:
SOURce on page 1649
EXPort:WAVeform:NAME on page 1651
EXPort:WAVeform:SAVE on page 1651
Mixed signal option (MSO, R&S R
T
O6-B1)

Table of Contents

Other manuals for R&S RTO6