EasyManua.ls Logo

R&S RTO6 - Page 2439

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2439User Manual 1801.6687.02 ─ 05
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Usage: Asynchronous command
TRIGger<m>:PARallel:PATTern:WIDTh:RANGe <WidthRangeMode>
Selects how the range of a pulse width is defined if TRIGger<m>:PARallel:
PATTern:
MODE is set to WIDTh.
Parameters:
<WidthRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers when the pattern comes false inside a given time
range. The time limit is defined by TRIGger<m>:
PARallel:
PATTern:WIDTh[:WIDTh] and TRIGger<m>:PARallel:
PATTern:WIDTh:DELTa (Width
±
Delta).
OUTSide
Triggers when the pattern comes false before or after the given
time range. The time limit definition is the same as for WITHin
range.
SHORter | LONGer
Triggers when the pattern comes false before or after the given
width has expired. Width is set with TRIGger<m>:PARallel:
PATTern:WIDTh[:WIDTh].
*RST: WITHin
Usage: Asynchronous command
TRIGger<m>:PARallel:PATTern:WIDTh[:WIDTh] <Width>
The effect depend on the setting of the
TRIGger<m>:PARallel:PATTern:WIDTh:
RANGecommand.
For the ranges SHOrter and LONGer, the width defines the maximum and minimum
time limit, respectively.
For the ranges WITHin and OUTSide, the width defines the center of a range which is
defined by the limits "±Delta".
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
Usage: Asynchronous command
Mixed signal option (MSO, R&S RT
O6-B1)

Table of Contents

Other manuals for R&S RTO6

Related product manuals