Remote control commands
R&S
®
RTO6
2568User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1..*
Selects the component. The DUT has suffix 1.
Parameters:
<InputPort1>
<InputPort2>
<OutputPort1>
<OutputPort2>
ONE |
TWO | THRee | FOUR
Port number that is assigned to the input or output.
*RST: ONE for input 1, TWO for output 1, THRee for
input 2, FOUR for output 4
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:SPONe <SParamFilePath>
DEEMbedding<m>:COMPonent<n>:SPTW
o <SParamFilePath>
Set the file name, file format and path of the S-parameter file, and load the S-parame-
ters.
See also: "S-Parameters, Filetype" on page 1208.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1..*
Selects the component. The DUT has suf
fix 1.
Parameters:
<SParamFilePath>
String with path and file name with extension s2p (2-port com-
ponent) or s4p (4-port component).
Example: See Chapter 24.3.6.1, "Adding cables to the measurement
setup", on page 1273.
Usage: Asynchronous command
24.24.4 Responses
DEEMbedding<m>:CADC?
Maximum attenuation or gain of the measurement setup before deembedding at DC.
This value is the expected basic attenuation or gain, which is corrected by deembed-
ding.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Deembedding (option R&S RT
O6-K121)