Acquisition and waveform setup
R&S
®
RTO6
171User Manual 1801.6687.02 ─ 05
5.7 R&S RT-ZVC probe
With the R&S
RT
O6 and option R&S RTO6-B1E, you can use the R&S RT-ZVC multi-
channel power probe. It has an integrated 2- or 4-channel amperemeter and 2- or 4-
channel voltmeter. The probe provides parallel measurements of analog or digital, volt-
age/current signals with excellent 18-bit resolution.
For more information on the R&S RT-ZVC probe, see also its user manual.
Source Channels
You can simultaneously connect a R&S RT-ZVC and a R&S RT-ZL04 to the
R&S RTO6, but no parallel operation on screen is possible.
You can acquire and measure the R&S RT-ZVC or R&S RT-ZL04 together with the
analog input channels. They are running on the same horizontal scale.
If an amperemeter or voltmeter channel is activated, it can be displayed on the screen
and used as a source for:
●
Cursor measurements
●
Automatic measurements, and also histogram, limit checks, longterm, track, quick
measurements
●
Mask testing
●
Basic and advanced maths
●
Reference waveform
●
T
rigger type "Edge"
●
Search: all search events
●
XY diagrams
●
Data export
All features of the base unit and the R&S RTO6-K37 option are supported.
Data export
You can save the data of the amperemeter and voltmeter channels to an XML, CSV, or
BIN file. One channel per file can be saved. Files in BIN format can be reloaded to the
R&S RTO6 as reference waveforms.
See also:
●
Chapter 12.2.7, "Saving and loading waveform data", on page 473
●
Chapter 12.2.1, "Waveform export files", on page 453
Remote commands for export to file:
●
EXPort:WAVeform:SOURce on page 1649
●
EXPort:WAVeform:NAME on page 1651
●
EXPort:WAVeform:SAVE on page 1651
R&S RT-ZVC probe