EasyManua.ls Logo

R&S RTO6 - 24.8.2 Time base

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1326User Manual 1801.6687.02 ─ 05
STOP
Stops the running acquistion.
Usage: Event
Asynchronous command
24.8.2 Time base
TIMebase:SCALe
.........................................................................................................1326
TIMebase:RANGe........................................................................................................1327
TIMebase:DIVisions?....................................................................................................1327
TIMebase:HORizontal:POSition.....................................................................................1327
TIMebase:REFerence...................................................................................................1327
TRIGger<m>:OFFSet:LIMited........................................................................................1328
AUToscale...................................................................................................................1328
ACQuire:POINts:AUTO.................................................................................................1328
ACQuire:POINts:AADJust.............................................................................................1329
ACQuire:POINts:MAXimum...........................................................................................1329
ACQuire:POINts:ARATe?..............................................................................................1329
ACQuire:SRATe........................................................................................................... 1329
ACQuire:SRReal..........................................................................................................1330
ACQuire:RESolution.....................................................................................................1330
ACQuire:POINts[:VALue].............................................................................................. 1330
TIMebase:ROLL:ENABle.............................................................................................. 1331
TIMebase:ROLL:STATe?...............................................................................................1331
TIMebase:ROLL:MTIMe................................................................................................1331
TIMebase:SCALe <TimeScale>
Sets the horizontal scale - the time per division on the x-axis - for all channel and math
waveforms.
The setting accuracy depends on the current resolution (sample rate).
No interpolation:
The resolution is an integer multiple of the ADC sample rate.
With interpolation:
Any value for the horizontal scale can be set.
Parameters:
<TimeScale> Range: 25E-12 to 10000 (RTO, RTP) | 5000 (RTE)
Increment: 1E-12
*RST: 10E-9
Default unit: s/div
Usage: Asynchronous command
Acquisition and setup

Table of Contents

Other manuals for R&S RTO6

Related product manuals