Remote control commands
R&S
®
RTO6
1328User Manual 1801.6687.02 ─ 05
TRIGger<m>:OFFSet:LIMited <State>
If ON, the horizontal position cannot be set outside the visible waveform diagram.
See also: TIMebase:
HORizontal:
POSition on page 1327
Suffix:
<m>
.
1..3
The numeric suffix is irrelevant.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
AUToscale
Performs an autoset process: analyzes the enabled channel signals, and obtains
appropriate horizontal, vertical, and trigger settings to display stable waveforms.
Rohde & Schwarz does not recommend using the autoset in remote control. To adjust
the oscilloscope remotely
, especially for automated testing applications, use the
remote commands that adjust the horizontal, vertical and trigger settings.”
Usage: Event
Asynchronous command
ACQuire:POINts:AUTO <ReclgthManual>
Selection to keep constant either the resolution or the record length when you adjust
the time scale (TIMebase:SCALe) or acquisition time (TIMebase:RANGe).
Parameters:
<ReclgthManual> RESolution | RECLength
RESolution
Resolution is kept constant. Set the required resolution value
with ACQuire:RESolution.
RECLength
The record length is kept constant. Set the required record
length value with ACQuire:POINts[:VALue].
*RST: RESolution
Usage: Asynchronous command