Protocol analysis
R&S
®
RTO6
760User Manual 1801.6687.02 ─ 05
Remote command:
BUS<m>:
RFFE:MINGap:
SELect on page 2059
BUS<m>:RFFE:MINGap:TIME on page 2059
13.17.2.3 Configuring RFFE signals
For details on configuration settings, see Chapter 13.17.2.1, "RFFE configuration set-
tings", on page 757.
1. Press the [Bus] key on the front panel.
2. Select the tab of the bus you want to set up, for example "SB1".
3. In the "Setup" tab, select the protocol: "RFFE".
4. Select the RFFE "Version".
5. Select the "Read mode".
6. Tap "SCLK", and select the waveform of the clock line.
7. Tap "SDATA", and select the waveform of the data line.
8. Set the logical thresholds.
9. If the signal has glitches which can distort the decoding, enable the glitch filter. Set
the glitch width.
10. Enable "Decode".
13.17.3 RFFE trigger
13.17.3.1 RFFE trigger settings
Access: [Bus] > "Shortcuts" tab > "Setup trigger"
RFFE (option R&S RT
O6-K550)