Remote control commands
R&S
®
RTO6
1392User Manual 1801.6687.02 ─ 05
Parameters:
<Level> Voltage for the trigger level.
Range: Depends on vertical scale, channel offset and other
settings. The trigger level must be within the current
display range.
Increment: 1E-3
*RST: 0
Default unit: V
Example:
TRIG:LEV5 0.01
Sets the trigger level for the external trigger signal to 10 mV.
TRIG2:
LEV3 0.2
Sets the trigger level for the B-event and B-trigger source chan-
nel 3 to 200 mV.
Usage: Asynchronous command
TRIGger<m>:FINDlevel
Sets the trigger level automatically. The command is only relevant if the trigger source
is an analog channel CHAN1...4.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Usage: Event
Asynchronous command
DISPlay:TRIGger:LINes <State>
Hides or shows the trigger levels in the diagrams.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
24.9.2 Edge trigger
TRIGger<m>:EDGE:SLOPe
..........................................................................................1392
TRIGger<m>:EDGE:SLOPe <Slope>
Defines the edge for the edge trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Trigger