Remote control commands
R&S
®
RTO6
1837User Manual 1801.6687.02 ─ 05
24.17.9 LIN (option R&S RT
O6-K520)
Configuration.......................................................................................................1837
T
rigger.................................................................................................................1839
Decode results.................................................................................................... 1845
LIN search settings............................................................................................. 1852
LIN search results............................................................................................... 1859
24.17.9.1 Configuration
BUS<m>:LIN:DATA:SOURce........................................................................................ 1837
BUS<m>:LIN:DATA:THReshold.....................................................................................1837
BUS<m>:LIN:TECHnology............................................................................................1838
BUS<m>:LIN:BITRate.................................................................................................. 1838
BUS<m>:LIN:POLarity..................................................................................................1838
BUS<m>:LIN:STANdard............................................................................................... 1839
BUS<m>:LIN:DATA:SOURce <DataSource>
Sets the waveform of the data line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 |
R1 | R2 | R3 |
R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
See Chapter 24.4.2, "Waveform parameter", on page 1284
*RST: C1W1
Usage: Asynchronous command
BUS<m>:LIN:DATA:THReshold <Threshold>
Sets a user-defined threshold value.
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:
LIN:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Threshold> Range: -12 to 12
Increment: 0.1
*RST: 0
Default unit: V
Protocols